Specialty Instruments · MT
Dew Point Transmitters — Compressed Air and Process Dew Point Measurement
Pioneer supplies dew point transmitters for compressed air, dryers and process gases, measuring -60 to +20 °C dew point (-80 to +20 °C available) with ±2 °C dew point accuracy and a 4–20 mA output. The stainless steel housing is rated to 5 MPa (15 MPa available) with a G1/2 connection.
- Dew point range: -60 to +20 °C td; -80 to +20 °C td available
- Accuracy: ±2 °C td
- Output signal: 4–20 mA (typical)

Overview
What it does and where it fits.
Dew point is the temperature at which water vapour in a gas starts to condense. Unlike relative humidity, it does not change with the gas temperature, which makes it the right measure of how dry compressed air or a process gas really is. A high dew point means wet air; a low dew point means dry air.
In compressed air systems, a dew point transmitter after the dryer confirms that the dryer is working and warns before moisture reaches tools, instruments, pneumatic valves or product. It is also used to control desiccant dryer regeneration, to monitor drying processes, and to protect moisture-sensitive products in pharmaceuticals, food and electronics.
Pioneer's transmitter is rated for line pressure up to 5 MPa (15 MPa available), so it can be installed directly in the compressed air line or in a sampling cell. The sensor is protected by a stainless steel woven filter with pores smaller than 8 µm.
Types
- Compressed air dew point transmitterIn-line 4–20 mA transmitter, -60 to +20 °C td, for dryer outlets and compressed air distribution.
- Low dew point versionExtended range down to -80 °C td for desiccant dryers and very dry process gases.
- High pressure versionRated to 15 MPa for high-pressure air and gas systems.
- Low ambient versionOperating temperature -30 to +50 °C for outdoor or unheated installations.
Features
- -60 to +20 °C td range; -80 to +20 °C td available
- ±2 °C td accuracy, 0.5 °C td repeatability
- 4–20 mA output
- Rated to 5 MPa; 15 MPa available
- G1/2 process connection, other connections on request
- Stainless steel housing with <8 µm woven SS sensor filter
- EMC according to IEC 61326-1
Applications
- Compressed air dryer outlets
- Desiccant dryer regeneration control
- Compressed air for instruments and pneumatic controls
- Pharmaceutical and food-grade compressed air
- Drying processes for plastics and materials
- Electronics manufacturing
- Condensation prevention in HVAC/R
Specifications
Spec table.
Standard ranges and materials. Tell us your medium, range and connection and we will match the model.
| Parameter | Value |
|---|---|
| Dew point range | -60 to +20 °C td; -80 to +20 °C td available |
| Accuracy | ±2 °C td |
| Repeatability | 0.5 °C td |
| Response time (to 63 %) | 5 s above -50 °C td; 5 to 300 s below -50 °C td |
| Output signal | 4–20 mA (typical) |
| Supply | 9 V DC (6–12 V DC) |
| Operating temperature | 0 to +50 °C; -30 to +50 °C available |
| Operating pressure | 5 MPa; 15 MPa available |
| Pressure connection | G1/2 (other connections on request) |
| Housing material | Stainless steel |
| Sensor protection | Stainless steel woven filter, pore size <8 µm |
| EMC | According to IEC 61326-1 |
Other dew point ranges, process connections and operating conditions on request.
For consultants
Consultant spec clause.
Paste it into your technical specification or BOQ, then adjust ranges and quantities to your schedule.
1. Dew point transmitters shall be provided at the outlet of each compressed air dryer and at the locations shown on the drawings. 2. Measuring range shall be -60 to +20 °C dew point [-80 to +20 °C for desiccant dryers with a design dew point below -60 °C]. 3. Accuracy shall be ±2 °C dew point or better, with repeatability of 0.5 °C dew point. 4. Output shall be 4–20 mA for connection to the compressor controller, PLC or BMS. 5. The transmitter shall be rated for a working pressure of 5 MPa [15 MPa] and shall have a stainless steel housing with a G1/2 process connection. 6. The sensor shall be protected by a stainless steel filter with pore size smaller than 8 µm, and the transmitter shall comply with IEC 61326-1 for EMC. 7. The transmitter shall be installed in the line or in a sampling cell with a controlled bleed, mounted from the top or side of the pipe and away from condensate traps and dead legs. 8. Each transmitter shall be supplied with a manufacturer's test/calibration certificate. Make: Pioneer Industrial Instruments, Pune, or approved equivalent.
What dew point do I need for compressed air?
It depends on what the air serves. Refrigerant dryers typically give a pressure dew point around +3 °C, which suits general plant air indoors. Instrument air, outdoor lines and pharmaceutical or food applications usually need desiccant dryers at -40 °C, and critical uses go to -70 °C. ISO 8573-1 defines these humidity classes. Pioneer's transmitter covers -60 to +20 °C td, with -80 °C available.
What is the difference between pressure dew point and atmospheric dew point?
Pressure dew point is the dew point at the actual line pressure, which is what matters for condensation in the pipework. When air is expanded to atmospheric pressure its dew point drops, so the same air reads much drier at atmosphere. Always specify and measure pressure dew point for compressed air. Pioneer's transmitter is rated to 5 MPa, or 15 MPa as an option, so it can measure at line pressure.
Why is dew point better than relative humidity for dry air?
Relative humidity depends on temperature: the same air shows a different %RH as it warms or cools. At very low moisture levels, RH readings also become too small to resolve. Dew point is an absolute measure of water vapour content and stays the same regardless of air temperature, so it shows clearly whether a dryer is meeting its specification.
How should a dew point transmitter be installed?
Install it after the dryer and filters, on the top or side of the pipe so liquid water and oil do not collect on the sensor. A sampling cell with a small controlled bleed gives faster response and lets the transmitter be removed without depressurising the line. Avoid dead legs. At very low dew points, below -50 °C td, response is slower, from 5 up to 300 s to 63 %.