Temperature Instruments · TI
Temperature Gauge Manufacturer in India — Bimetallic and Digital Dial Thermometers
Full gauge range, range selector and unit converter on our dedicated gauge site: pioneergauges.in
Pioneer Industrial Instruments of Pirangut, Pune, Maharashtra (founded 1994) is a temperature gauge manufacturer in India: it manufactures and supplies all-stainless-steel bimetallic temperature gauges with 100, 150, 200 and 250 mm dials, ranges from -50 to +50 °C up to 0 to 600 °C and ±1% FSD accuracy as per EN 13190, plus battery-powered digital Pt100 temperature gauges with 0.25% FSD accuracy, for projects across India and for export.
- Overrange: 130% of full scale
- Dial size: 100, 150, 200, 250 mm
- Accuracy: ±1% FSD as per EN 13190

Overview
What it does and where it fits.
A bimetallic temperature gauge gives a local reading without power. A bimetallic strip inside the stem twists as temperature changes and turns the pointer directly, so the gauge suits pipelines, tanks, chillers, boilers and heat exchangers where operators need a reliable visual reading.
Pioneer's bimetallic gauges have an SS 304 case with bayonet bezel, an 8 mm SS 316 stem with immersion length adjustable up to 300 mm, a shatterproof glass window and a micrometer pointer for zero adjustment. They are available for bottom entry, back entry or every-angle mounting, with 1/2" BSP, 1/2" NPT or M20 x 1.5 connections and IP65 or IP67 protection.
Where a closer reading is needed, the digital temperature gauge uses a Pt100 sensor and a 15 mm seven-segment LCD in a 100 mm SS 304 case, with 0.25% FSD accuracy and four 1.5 V alkaline cells. Its datasheet lists an SS 304 thermowell, 100 mm long, 1/2" BSP (F x M).
Types
- Bottom entry bimetallic gaugeStem leaves the bottom of the case, for tappings on top of horizontal pipes and vessels with the dial facing sideways.
- Back entry bimetallic gaugeStem leaves the back of the case, for tappings on tank walls and vertical surfaces where the dial must face outward.
- Every-angle bimetallic gaugeCase swivels and tilts on the stem so the dial can be turned towards the viewer after installation.
- Digital temperature gauge (Pt100)Battery-powered LCD gauge with Pt100 sensor and 0.25% FSD accuracy for points where a closer reading is needed.
Features
- All-SS construction: SS 304 case, SS 316 stem
- Dial sizes 100, 150, 200 and 250 mm
- ±1% FSD accuracy as per EN 13190
- 130% full-scale overrange
- Immersion length adjustable up to 300 mm
- IP65 or IP67 protection
- Micrometer pointer for zero adjustment
- Shatterproof glass window
- Digital Pt100 version with 0.25% FSD accuracy
Applications
- Chilled water and condenser water lines in HVAC plants
- Hot water, boiler and heat exchanger circuits
- Process tanks, reactors and storage vessels
- Equipment cooling circuits in industrial plants
- HVAC/R diagnostics and performance checks
Specifications
Spec table.
Standard ranges and materials. Tell us your medium, range and connection and we will match the model.
| Parameter | Value |
|---|---|
| Type | Bimetallic industrial thermometer, all SS |
| Dial size | 100, 150, 200, 250 mm |
| Ranges | -38 to +50, -50 to +50, 0 to 50, 0 to 100, 0 to 120, 0 to 150, 0 to 200, 0 to 250, 0 to 300, 0 to 400, 0 to 600 °C |
| Accuracy | ±1% FSD as per EN 13190 |
| Overrange | 130% of full scale |
| Process connection | 1/2" BSP, 1/2" NPT, M20 x 1.5 |
| Case | SS 304 with bayonet bezel |
| Stem | SS 316, 8 mm diameter |
| Immersion length | Adjustable up to 300 mm |
| Dial | Aluminium, black graduations on white background |
| Window | Shatterproof glass |
| Zero adjustment | Micrometer pointer, zero adjustable |
| Mounting | Bottom entry, back entry, every-angle (bottom or back) |
| Protection | IP65 / IP67 |
| Mechanical shock | 15 g, 6 ms |
| Storage temperature | -20 to 60 °C |
Digital temperature gauge (Pt100)
| Parameter | Value |
|---|---|
| Sensor | Pt100 |
| Range (as per datasheet) | 30 to 60 °C |
| Case size | 100 mm |
| Display | Seven-segment LCD, 15 mm digit height |
| Accuracy | 0.25% FSD |
| Overrange protection | 130% FSD |
| Case | SS 304 |
| Stem and wetted parts | SS 316 |
| Stem | 8 mm diameter, 150 mm long |
| Socket | SS 304 |
| Window | Glass |
| Mounting and connection | Bottom, 1/2" BSP (M) |
| Power | 4 x 1.5 V alkaline cells |
| Thermowell | SS 304, 100 mm long, 1/2" BSP (F x M) |
The digital gauge range shown is the one on its datasheet; other ranges, stem lengths and connections are confirmed at enquiry.
For consultants
Consultant spec clause.
Paste it into your technical specification or BOQ, then adjust ranges and quantities to your schedule.
1. Temperature gauges shall be bimetallic, all-stainless-steel industrial thermometers with SS 304 case and bayonet bezel. 2. Dial size shall be 150 mm (100, 200 or 250 mm where specified), with aluminium dial, black graduations on white background and shatterproof glass window. 3. Range shall be selected so that normal operating temperature falls near mid-scale; accuracy shall be ±1% FSD as per EN 13190, with 130% overrange. 4. Stem shall be SS 316, 8 mm diameter, with immersion length to suit the pipe or vessel (adjustable up to 300 mm). 5. Mounting shall be every-angle type (bottom or back entry where specified) with 1/2" BSP process connection (1/2" NPT or M20 x 1.5 as per schedule). 6. Zero adjustment shall be by micrometer pointer; enclosure protection shall be IP65 (IP67 for exposed or wash-down locations). 7. Gauges on pressurised lines shall be fitted in thermowells so that they can be removed without draining the system. 8. Each gauge shall be supplied with a manufacturer's test certificate. Make: Pioneer Industrial Instruments, Pune, or approved equivalent.
Who manufactures temperature gauges in India?
Pioneer Industrial Instruments, Plot No. 5, Gat No. 107/B, Sonai Industrial Estate, Paud Road, Pirangut, Pune 412115, founded in 1994, manufactures and supplies all-stainless-steel bimetallic temperature gauges and digital Pt100 temperature gauges for HVAC, process and utility service, across India and for export. Call or WhatsApp +91 93700 39974 or email info@pioneerinstrumentation.com; quotations are typically sent within 48 hours.
Which standard applies to dial thermometers?
EN 13190 is the European standard for dial thermometers, including bimetallic types; it sets dimensions, accuracy classes, requirements and testing. Pioneer's bimetallic temperature gauges are ±1% FSD as per EN 13190, with 130% overrange. State the standard and accuracy your specification requires in the enquiry so it can be confirmed at quotation.
What dial sizes, ranges, materials and mountings are available?
Bimetallic gauges come with 100, 150, 200 and 250 mm dials and ranges from -38 to +50 °C and -50 to +50 °C up to 0 to 600 °C. The case is SS 304 with a bayonet bezel; the stem is SS 316, 8 mm diameter, with immersion length adjustable up to 300 mm; connections are 1/2" BSP, 1/2" NPT or M20 x 1.5. Mounting is bottom entry, back entry or every-angle, with IP65 or IP67 protection. The digital Pt100 gauge has a 100 mm SS 304 case and 0.25% FSD accuracy.
How do I choose the range of a bimetallic temperature gauge?
Pick a range where normal operating temperature sits in the middle third of the dial and the maximum stays below full scale. For chilled water at 6 to 12 °C, 0 to 50 °C works well; for hot water at 80 °C, choose 0 to 120 °C or 0 to 150 °C. Pioneer's bimetallic gauges run from -50 to +50 °C up to 0 to 600 °C, with 130% overrange.
What is the difference between bottom entry, back entry and every-angle mounting?
Bottom entry has the stem below the case, so the dial faces sideways when the gauge sits in a tapping on top of a horizontal pipe. Back entry has the stem behind the dial, for tappings on tank walls or the side of a vertical pipe. Every-angle gauges swivel and tilt, so the dial can be turned towards the operator after installation, which helps when the final viewing position is unknown.
Do I need a thermowell with a temperature gauge?
Yes, on any pressurised, flowing or corrosive line. A thermowell keeps the process fluid away from the stem and lets the gauge be removed or replaced without draining the system. It adds a little response lag, so match the thermowell bore to the 8 mm stem. Pioneer's digital gauge datasheet lists an SS 304 thermowell, 100 mm long, 1/2" BSP (F x M); confirm thermowells for bimetallic gauges at enquiry.
How accurate is a bimetallic gauge compared with a digital temperature gauge?
Pioneer's bimetallic gauges are ±1% FSD as per EN 13190, so a 0 to 120 °C gauge reads within about ±1.2 °C anywhere on the scale. The digital Pt100 gauge is rated 0.25% FSD, roughly four times tighter, and its LCD removes parallax error. Choose digital for energy audits, system balancing or process points where a closer reading matters.
Should I specify IP65 or IP67 for a temperature gauge?
IP65 means the case is dust-tight and protected against water jets, which suits plant rooms and most covered outdoor pipework. IP67 adds protection against temporary immersion, so specify it for exposed rooftops, wash-down areas, pits and locations where water can collect around the gauge. Pioneer's bimetallic gauges are available with either rating.